Become a Partner

Become a Partner

Is your chlorinator showing 0 ppm? It might be in standby mode. Follow this guide to troubleshoot power settings, flow issues, and low salt readings.

Is your salt chlorinator showing 0 ppm, a very low salt reading, or no chlorine production? Don't panic. In many cases, the unit is simply in "Standby Mode" or needs a quick flow adjustment.

Category Possible Cause Symptoms How to Check Solution
Unit Not Fully Powered On System is in standby mode (not actually running) App shows 0 ppm; no chlorine production; power light not green In the app, power icon is white (not blue); power light on unit is red (not green) In the app, tap the power button (turns blue) and press the physical power button on the unit (turns green). Device only starts detecting and displaying salt readings when powered on.
No or Low Water Flow Detected Pump speed too low / valve not fully open / dirty filter Flow indicator light flashing; no salt reading Confirm pump is running; check flow indicator light status Increase water flow; clean filter; open valves fully; meet minimum flow rate: Horizontal ≥60 L/min, Vertical ≥90 L/min.
Flow Switch Blocked by Debris Flow paddle or impeller blocked by branches, leaves, debris Flow indicator light flashing; salt reading not updating Check if flow switch paddle moves freely Remove debris (leaves, scale, hair, etc.); ensure paddle moves freely; re-test.
Incorrect Installation Direction or Orientation Inlet/outlet reversed / vertical installation with low flow Flow light flashing; no salt reading Check body arrow direction; confirm mounting orientation Reinstall in correct arrow direction; horizontal installation preferred; ensure adequate flow.
Salt Recently Added, Not Fully Dissolved Salt reading delay after addition Shows 0 ppm or unstable readings after adding salt Look for undissolved salt crystals in water; confirm time since addition Run pump continuously for several hours (typically ~24 hrs) before re-checking.
Actual Salt Level Too Low Not enough salt in water Consistently low or 0 ppm Test with salt test kit or strips Add salt to recommended level (e.g., 3000–3500 ppm) and circulate evenly.
Startup/Reading Delay System just started up Short-term 0 ppm after powering on Observe for a few minutes after startup Wait a few minutes; device displays reading after detecting water flow.
Cell Scaling Calcium scale buildup on plates reduces conductivity Low or inaccurate salt reading; reduced chlorine output Visually inspect cell plates Clean cell; set polarity reversal time to 4 hours to reduce scaling.
Test Strips Misleading Oxidizers/chlorine affect strip color Test strip reading higher than device Compare strip and device readings Use device reading as main reference; test strips may show higher values.
E2 (Low Salt) Status Controller detects low salt E2 displayed; unsure if chlorine is being produced Check code + observe chlorine production In E2 state, unit still produces chlorine at about 50% output (reduced but not stopped).

Use this step-by-step guide to diagnose and fix the issue quickly.

1. The Most Common Cause: Unit Not Fully Powered On

Symptom: App shows 0 ppm; no chlorine production; power light is NOT green.

The Fix: Many users mistake "Standby Mode" for "On."

  • Check the App: Look at the power icon. If it is White, the system is in standby. Tap it until it turns Blue.

  • Check the Unit: Press the physical power button on the controller until the light turns Green.

  • Note: The device only starts detecting and displaying salt readings when fully powered on.

2. Water Flow Issues

Symptom: Flow indicator light is flashing; salt reading is not updating or shows 0.

Possible Causes:

  • Low Flow: Pump speed is too low or valves are not fully open.

  • Blockage: Debris (leaves, branches, hair) is blocking the flow paddle.

  • Wrong Direction: The cell is installed backwards (inlet/outlet reversed).

The Fix:

  • Increase Flow: Ensure flow rates meet the minimum: Horizontal ≥60 L/min, Vertical ≥90 L/min.

  • Clean Debris: Remove the cell and check if the flow switch paddle moves freely. Remove any trapped debris.

  • Check Orientation: Ensure the arrow on the cell body matches the water flow direction.

3. Salt Not Fully Dissolved

Symptom: You just added salt, but the reading is still low or unstable (0 ppm).

The Fix:

  • There is a delay between adding salt and the sensor reading it. Undissolved salt crystals sitting on the bottom won't register immediately.

  • Wait: Run your pump continuously for 24 hours to allow salt to fully dissolve and mix before re-checking the reading.

4. Cell Scaling (Calcium Buildup)

Symptom: Low or inaccurate salt reading; visible white scale on plates.

The Fix:

  • Inspect: Visually inspect the titanium plates inside the cell.

  • Clean: If you see white calcium buildup, clean the cell with a mild acid solution (as per manual).

  • Prevent: Set the "Polarity Reversal" time to 4 hours in the settings to improve self-cleaning performance.

5. Understanding the "E2" (Low Salt) Status

Symptom: The controller displays "E2".

What it means: The controller detects low salinity. However, in an E2 state, the unit does NOT stop working. It continues to produce chlorine at a reduced rate (approximately 50% output) to protect the cell.

The Fix: Test your water with a separate salt test kit. If salt is truly low, add salt to reach the recommended 3000–3500 ppm range.

Quick Summary Checklist

If you see an abnormal reading, do this first:

  1. Check Power: Ensure App icon is Blue and Unit light is Green.

  2. Check Flow: Ensure the Flow light is steady (not flashing).

  3. Check Debris: Ensure the flow paddle moves freely.

  4. Wait: If you just added salt, wait 24 hours.

Real Customer Case: We found that 90% of "No Reading" tickets were resolved simply by pressing the power button to exit standby mode, or clearing a small leaf stuck in the flow switch.

Inside TROXLLY

Installation & Buying Guides